OPERATING INSTRUCTIONSOAKTON®35618-SeriespH 300 and 310Portable Waterproof pH/mV/°C MeterR1 1/01pH/mV/°C Data MeterpH 300 SeriesWATERPROOFMODEHOLDENTE
1918IMPORTANT: For manual compensation,you must disconnect the temperature probe(see page 9).1. Switch the meter on. Press the MODEkey to select pH mo
To take readings:1. Rinse the probe with deionized or distilled water before use to remove anyimpurities adhering to the probe body. If the pH electro
This feature lets you freeze the value of the pH, mV (or Relative mV) and temperature readings for a delayed observation. HOLD can be used any time wh
25248. Advanced set up functionsSETUPMEMpHSETUPpHSETUPSETUPMEASSETUPSETUPMEMpHSETUPpHSETUPSETUPSETUPMEASSETUPThe advanced set up mode lets you customi
pH 300 meter set up mode2726Advanced set-up mode overviewP3.0: Viewing electrode dataIn pH measurement mode:• P3.1: pH electrode offset• P3.2: pH elec
From measurement mode: 1. Press the Set up key to enter Set Upmode.See figure2. Press the ▲ and ▼ keys to scroll throughsubgroups until you view param
P3.1: Electrode offsetFrom measurement mode1. Press the Setup key to enter Set Upmode.2. Press the ▲ and ▼ keys to scrollthrough subgroups until you v
From measurement mode1. Press Setup key to enter Set Up mode.2. Press the ▲ and ▼ keys to scroll throughsubgroups until you view parameter P4.0. See f
3534P4.3 Calibration buffer selection setsAvailable in pH 310 meter onlyThe pH 310 meter lets you select between three standard calibration buffer set
3736P5.0: Resetting to factory default settingsAvailable in pH 300 meter onlyIn the pH 300 meter, Program 5 lets you reset all parameters to factory d
32Table of Contents1. Introduction...42. Display and keypad f
3938P6.0: Resetting to factory default settingsAvailable in pH 310 meter onlyIn the pH 310 meter, Program 6 lets you reset all parameters to factory d
4140pH electrode activation (for glass bodied electrodes only)WARNING: Only qualified persons proficient with the safe handling of dangerouschemicals
434211. Error MessagesLCDIndicates Cause SolutionDisplayErr Unrecognized input Wrong input in Release key. Select annunciator from keypad selected mo
454413. AccessoriesReplacement meters35618-02 Replacement waterproof pH 300 meter35618-70 Waterproof pH 300 meter kit. Includes “All-in-One” probe, fi
474614. WarrantyOAKTON warrants this meter to be free from significant deviations in material and workmanship for a period of three years from date of
541. IntroductionThank you for selecting an OAKTON meter. This OAKTON portable meter is a microprocessor-based instrument that measures pH, mV, and te
76KeypadThe large membrane keypad makes the instrument easy to use. Each button, whenpressed, has a corresponding graphic indicator on the LCD.ON/OFF.
98Connecting the Electrode and Temperature ProbeThe OAKTON pH/mV meter can use any standard pH, ORP, or ISE electrode with aBNC connector. For automat
11103.3Attaching the Electrode Holderto the MeterAttaching the electrode holder to the meter facilitates one-hand operation. Attach two electrode hold
1312pH 300 meter calibrationThe pH 300 meter is capable of up to 5-point pH calibration to ensure accuracy acrossthe entire pH range of the meter. Sel
1. If necessary, press the MODE key toselect pH mode. The pH indicator appears in the upper right hand corner of the display.See figure 2. Rinse the p
Temperature calibration1. Make sure the ATC probe (or temperature connector of the “All-in-One” electrode) is attached to the 6-pinconnector. The ATC